SE-12 Series Electrocardiograph User Manual
Input Impedance:
Input Circuit Current:
Input Voltage Range
Calibration Voltage:
DC Offset Voltage:
Minimum Amplitude:
Noise:
Multichannel crosstalk
Filter
CMRR
Pacemaker Detection
Amplitude
Width
External Input/Output
Input
Output
WIFI (Configurable)
Transmitting Frequency
Frequency Band
Wireless protocol
≥100MΩ (10Hz)
≤0.01μA
≤± 5 mVpp
1mV± 2%
± 900mV
20 μVp-p
≤12.5 μVp-p
≤0.5mm
AC Filter: On/Off
DFT Filter:
0.01Hz/0.05Hz/0.15Hz/0.25Hz/0.32Hz/0.5Hz/0.67Hz
EMG Filter: Off/25Hz/35Hz/45Hz
LOWPASS Filter: 350Hz/300Hz/270Hz/150Hz/100Hz/75Hz
≥140dB (AC ON)
≥123dB (AC OFF)
± 500uV to ± 700 mV
30μs to 2.0 ms
≥100kΩ; Sensitivity 10mm/V±5%;
Single ended
≤100Ω; Sensitivity 1V/mV±5%;
Single ended
2400-2497MHz
2400-2497MHz
IEEE 802.11b/g/n
- 96 -