2 Installation
• Refer to the electrical connections below before finally fitting the instrument into the equipment
rack.
• Temporarily locate and screw in the two spigots* to each side of the equipment rack.
Locate and slide the instrument into the rack, locating the instrument on the spigots*.
• Secure the instrument in the equipment rack with two of the screws and washers (supplied).
• Remove the two spigots* and replace with the remaining two screws and washers (supplied).
2.6

Electrical connections

WARNINGS
1.
THE GROUND LEAD OF THE INSTRUMENT MUST BE CONNECTED TO THE AC SUPPLY
PROTECTIVE SAFETY GROUND.
2.
ISOLATE THE POWER SUPPLY BEFORE MAKING ANY ELECTRICAL CONNECTIONS TO THE
REAR PANEL.
General
The instrument must be connected to the correct electrical power supply as stated on a label next
to the power connector. Also see section 6 Reference and Specification.
Make sure that the power supply is off before connecting the power cable.
Requirements for rack-mounted instruments
• Install an isolator in the power supply circuit. The power supply connector and switch on the
rear panel of the instrument will not be accessible.
• Set the power supply isolator to OFF. Connect the power supply and set the power supply
switch to ON before sliding the instrument into the rack.
• Set the power supply isolator to ON .
Check that the front panel display shows the power-up sequence.
Connecting (Figure 2-6)
To connect the power supply to the instrument proceed as follows:
• Insert the IEC power connector (1) into the power supply connector at the rear of the
instrument.
• Set the ON/OFF switch to ON.
• Check that the front panel display shows the power-up sequence.
K0443 Issue No. 3
2 - 8