4 Maintenance
•
Reconnect the electrical power supply connector.
Note:
For rack-mounted units, set the power switch to on, locate and secure the
instrument in the rack.
•
Apply the pneumatic pressure and/or vacuum supplies - switch on the electrical
power supply.
•
Switch on the instrument and carry out a serviceability test detailed in section 5.
PACE 6000
Cautions:
1.
AFTER REMOVING A CONTROL MODULE
2.
-
TWO
CHANNEL OPERATION
BE DONE WHEN BOTH PRESSURE MODULES ARE BELOW
DAMAGE OCCURS IF ONE PRESSURE MODULE IS BELOW
•
When installing two pressure modules, the module with the higher pressure
range must be located in the right-hand side (Module 1) above the power supply
connector.
•
If using two pressure modules of the
same pressure range, the module with
the lower serial number must be located
in the right-hand side (Module 1) above
the power supply connector.
Note:
When connecting the output ports of two pressure
modules together make sure both are either:
• below 70 bar.
OR
• between 100 to 210 bar.
To comply with the Pressure Equipment Directive (PED) do not mix these categories.
K0443 Issue No. 3
,
USE A BLANKING PLATE TO KEEP THE FLOW OF COOLING AIR
,
COMBINING THE OUTPUTS OF TWO PRESSURE CONTROL MODULES CAN
4 - 6
70
100
BAR OR BETWEEN
70
BAR AND THE OTHER IS
.
210
.
TO
BAR
100
210
.
TO
BAR
Serial number