PACE Pressure Controller User Manual
IEEE 488
Located on the rear panel an external IEEE 488 connection requires:
Connector
Communications
Default Address
Protocols †
New data up-date rate
Built-in EMC filtering and transient voltage protection.
Timeout
Presets the times for automatically changing from control to measure mode.
Note:
Controller timeouts can save supply gas, extending control valve life and
minimising acoustic noise.
Idle Timeout starts when the controller achieves the set-point after the set time, if new set-points
are not entered the timer will time-out and return to measure mode.
Gas Head Correction
Corrects pressure output for the height difference between
instrument and UUT:
for UUT positioned higher than the reference level
of the PACE instrument enter a positive height
correction.
for UUT positioned lower than the reference level of
the PACE instrument enter a negative height
correction.
When calibrating the PACE instrument disable the gas head correction and correct the actual
applied pressures for height.
K0443 Issue No. 3
=
24-way 'D' female wired as IEEE 488 standard
=
IEEE 488 GPIB
=
16
=
SCPI, DPI 500, DPI 510, DPI 515, DPI 520
=
refer to data sheet
† Not all protocols available on all models.
6 - 13