PACE Pressure Controller User Manual
Divider
Select and set-up the divider task by pressing Divider from the task screen. The divider menu
defines high set-point, low set-point and then divides the span into a number of equal test points
(min 2, max 25).
Divider menu structure
Select required units, Rate, etc. in
the set-up menus. When Divider is
then entered from the Task menu,
these test point pressures and
number of test points can be set.
By entering control mode, enables
the Divider sequence of test
pressures (and controlled at the
selected rate).
Example:
High set-point = 2 bar
Low set-point = 0 bar
Number of points = 5
Test pressures =
0, 0.5, 1, 1.5 and 2 bar
1
3
4 4
2
P = pressureN = Number of pointsR = range between low set-point (1) and high set-point (5)
*
controlled pressure to each set-point
K0443 Issue No. 3
3 - 11