PACE Pressure Controller User Manual
Key to Figure 2-3
1
Pressure source
4
Regulate to between 110% full-scale and MWP
5
Diffuser*
8
Protection device
10
Oil mist trap
13
Check valve ‡
Notes:Refer to Section 6 Reference and Specification for details of other system components.
*
High pressure gas exhaust - depending on pressure range.
†
Optimum controller transient response and minimum time to set-point may be degraded if either
the pneumatic supply or vacuum system has restricted flow. Installing a reservoir volume, which
has larger capacity than the load volume, located in close proximity to the controller supply ports
can improve the controller response.
‡
Optional vacuum system kit.
8
To protect the control module, for ranges above 100 bar, from over-pressure a suitable protection
device (such as a relief valve or bursting disc) must be fitted to limit the applied supply pressure to
below the MWP.
Optional differential connection kit.
K0443 Issue No. 3
12
11
Figure 2-3, Pneumatic Connections with vacuum supply
2
Conditioner
6
Unit under test
9
Optional differential connection
8
11
Vacuum source
a
atmosphere
SUPPLY
3
7
12
2 - 5
OUTLET
VENT REF
13
*
Filter
Optional reservoir †
Normally open electrical release valve
a