User Guide
3.4.1

RECORD SETUP

Select Record Setup to define the following settings:
• Patient group:
• Duration:
• Deflation speed:
3.4.2

Program

Selecting a Program
Enter program setup and select Change to select the program (A, B,
C or D). The recording program defined is used when a long term
record is started.
Defining the Recording Programs
Enter the program setup start times, sleep times and duration for the
four programs.
Highlight the recording program to be set and set the start time and
measurement duration for each entry. Up to 8 separate start times can
be defined with the BP measurement interval defined for each start
time.
1. Define the start time by pressing change. The time counts up in
10 minute intervals.
2. When the desired start time is shown press next. The
measurement interval is highlighted.
3. Press change to select a BP measurement interval of: 5, 10, 15,
20, 30, 60, or 120 minutes.
4. Press next to define the next start time.
5. Continue until all desired start times (with measurement interval)
are defined for the selected program.
When less than 8 starts are required pressing next twice returns to the
previous menu. A minimum of 2 start times must be defined.
Long Term Recording
adult or child
24 hours or 48 hours
2, 3, 4, 5, 6, 7, 8, 9 mmHg/s, or Auto. The
Auto setting sets the deflation speed ac-
cording the heart rate.