Schiller BR-102 PLUS User Manual page 59

24/48 hour ambulatory blood pressure recorder
Table of Contents

Advertisement

BR-102 plus / BR-102 plus PWA
No.
Message
182
Signal
disturbed
190
No signal
191
Weak signal
192
Signal
disturbed
198
DIA
not
detected
199
SYS
not
detected
200
No pulse
201
Pulse > max.
202
Pulse < min.
210
No pulse
211
Pulse > max.
212
Pulse < min.
238
Min. pressure
239
Max. pressure • Maximum pressure reached
255
Internal error
322
No
measurements
available
323
No patient data
available
324
SD-
card
missing
User Guide
Cause
• Too much interference in Ko-
rotkoff signal.
• Pressure reached 50 mmHg
and no pulse detected (OSC).
• Less than 8 pulse beats de-
tected (OSC).
• More than 200 pulse beats
detected (OSC).
• Minimum pressure reached
without diastolic value detect-
ed (OSC).
• Minimum pressure reached
without systolic value detect-
ed (OSC).
• No pulse detected (OSC).
• Too many pulse beats detect-
ed (OSC).
• Too few pulse beats detected
(OSC).
• No pulse detected (KOR).
• Too many pulse beats detect-
ed (KOR).
• Too few pulse beats detected
(KOR).
• Minimum pressure reached
without measurement result.
without measurement result.
• Device is defective.
• No recording stored.
• No recording has been made.
• Incorrect initialisation of a re-
cording.
• Error in the recording storage.
• No patient data has been re-
corded.
• SD card not inserted. For
PWA recording an SD card is
necessary.
Remedy
Perform measurement in quiet environment;
avoid moving the arm during measurement.
Check cuff.
Check cuff.
Avoid moving the arm during measurement.
Avoid moving the arm during measurement.
Avoid moving the arm during measurement.
Repeat measurement; check cuff.
Perform measurement in quiet environment;
avoid moving the arm during measurement.
Repeat measurement; check cuff.
Connect microphone; Check position of
microphone. Ensure that the microphone is
correctly orientated (metallic (yellow) side
towards arm).
Perform measurement in quiet environment;
avoid moving the arm during measurement.
Place microphone over brachial artery;
ensure that the microphone is correctly
orientated (metallic (yellow) side towards
arm).
Repeat measurement.
Repeat measurement.
Contact technical service.
Take/ retake a recording.
Enter patient data.
Insert an SD card.