7
Technical Data
7.1

Preventing electromagnetic interferences

"Non-ionic electromagnetic radiation"
HF source
Wireless communications devices
Various radio services (TETRA 400)
- Walkie-talkies (FRS)
- Rescue service, police, fire brigade, servicing (GMRS)
LTE band 13/17
- GSM800/900
- LTE band 5
- Radio telephone (microcellular) CT1+, CT2, CT3
- GSM1800/1900
- DECT (radio telephone)
- LTE Band 1/3/4/25
- UMTS
- Bluetooth, WLAN 802.11b/g/n
- LTE Band 7
- RFID 2450 (active and passive transponders and reading de-
vices)
WLAN 802.11a/n
Page 62
7.1
Preventing electromagnetic interferences
The user can help avoid electromagnetic disturbances by keeping the
minimum
distance
telecommunication devices (transmitters) and the BR-102 plus / BR-
102 plus PWA. The distance depends on the output performance of
the communication device as indicated below.
Portable HF telecommunication devices must not be used within
a radius of 0.3 m from the BR-102 plus / BR-102 plus PWA and its
cables.
Do not place the BR-102 plus / BR-102 plus PWA on top of other
electric/electronic devices - i.e. maintain a sufficient distance to
other devices (this includes the patient cables).
For permanent HF telecommunication devices (e.g. radio and TV), the
recommended distance can be calculated using the following formula:
. (The formula is based on the max. immunity level of 10 V/
d
=
0.6
P
m in the frequency domain of 80 MHz to 3000 MHz).
d = recommended minimum distance in meters
P = transmitting power in Watts
For more information on operation in an electromagnetic environment
according to IEC/EN 60601-1-2, please consult the service manual.
BR-102 plus / BR-102 plus PWA
between
portable
Transmitter fre-
Testing fre-
quency
quency
[MHz]
[MHz]
385
380-390
430-470
450
704-707
710/745/780
800-960
810/870/930
1720/1845/
1700-1990
1970
2400-2570
2450
5240/5500/
5100-5800
5785
and
mobile
HF
Max.
Distance d
Power P [W]
[m]
1.8
0.3
2
0.3
0.2
0.3
2
0.3
2
0.3
2
0.3
0.2
0.3