Schiller BR-102 PLUS User Manual page 37

24/48 hour ambulatory blood pressure recorder
Table of Contents

Advertisement

BR-102 plus / BR-102 plus PWA
User Guide
9. Check that the initial measurement is successful.
Initial Measurement Check
After the first measurement (on the resting patient), it is recommended
that the following is checked to help ensure that the recording is
successful:
12.02.2014
KOR
128
SYS
mmHg
DIA
• Ensure that a BP reading has been obtained and is displayed on the
BR-102 plus / BR-102 plus PWA.
• Check that the measurement has been taken using the auscultatory
method (KOR displayed next to the measurement, below the time).
Note: If the cuff does not have a microphone (or an unsuccessful
auscultatory measurement has been taken) OSC is displayed.
• If the measurement has not been taken correctly check the cuff
placement and microphone position.
• Check the battery display and ensure that it still shows full capacity.
10. After checking the first measurement and battery display, position
the BR-102 plus / BR-102 plus PWA in the pouch and secure.
– Subsequent measurements are taken as defined for the program
selected.
– During BP measurement, the ascending/descending cuff pres-
sure is shown. After a measurement has been taken, the result is
displayed for approximately one minute
cording on the BR-102 plus, page
Long Term Recording
10:19
MA
99
mmHg
PUL
86
71
bpm
OK
39).
BP recording
3
3.4
(see Displaying a Re-
Page 37