Introduction; Requirements; General; Equipment - VWR Mega Star 600R Service Manual

Hide thumbs Also See for Mega Star 600R:
Table of Contents

Advertisement

Introduction

This chapter describes the calibration work to be performed on a yearly basis.
A calibration certification form can be found in the attachment B-1.

General

Successfully completing the preventive maintenance checklist and calibration procedure as outlined
will ensure that the instrument is properly maintained and calibrated to VWR specifications.

Requirements

Equipment

Optical (Laser) tachometer
Thermocouple
Stopwatch
Weights according to rotor table
Safety Tester
Note: Check calibration due date on all test equipment before starting the procedure.
Documentation
Service manual
Centrifuge operating manual
Preventive maintenance checklist (A-1)
Calibration certification form (B-1)
Certification label
Certificate of calibration for each piece of test equipment used to perform the calibration
Training certificate
Preventive maintenance
Perform preventive maintenance checks as outlined in the service manual to ensure the
instrument is in good working order before performing any calibrations.
Complete the preventive maintenance checklist. (A-1)
CALIBRATION (REFERENCE MEASUREMENT)
6-2