Specifications
2.4.

Safety standards and EMC guidelines

The MultySonic 8000 has been developed and designed according to with the
following guidelines and standards.
Low-voltage guideline
EN 61010-1
IEC 255_5
Electromagnetic compatibility
EN 50082-2 (1995)
IEC 77A
EN 61000-4-11
EN 61000-4-4
EN 61000-4-5
EN 61000-4-12
IEC 1000-4-6
EN 61000-4-2
EN 61000-4-8
EN 61000-4-10
IEC 1000-4-3
Electro-magnetic emission
IEC 1000-3-2
EN 50081-2 (1994)
Ex protection certificates
2.5.

Interface specifications

2.5.1.

Analogue inputs

8x 4-20 mA standardised signals can be connected to the analogue inputs.
The potential difference of the signal terminals to the device earth must not
exceed 20 V.
The I/O board provides a maximum of 24 V max. 1 for passive water level
sensors.
The I/O board's inputs and outputs can be switched actively or passively
with microswitches (see the chapter 4.5).
Input signals smaller than the start of the measurement range and larger
than 21mA are evaluated as defective.
2.5.2.

Analogue outputs

The assigned variable is represented by a 4 -20 mA standard signal at the
active analogue outputs. The output can be assigned to the outflow, the
mean flow rate, the water temperature or various variables dependent on it.
The measuring ranges can be freely configured by entering the full
measuring scale.
If a variable is evaluated as defective, the analogue signal is reset to <3.6
mA. For values outside of the full measuring scale, the output takes the
relevant extreme value (3.84 or 20.5mA).
Safety Requirements for Electrical Equipment
Insulation: 2 kV common mode, 1 kV Normal mode, >100 MΩ
Part 2 Industrial
Harmonics on power supply
Interruptions (100ms), dips and voltage variations +12 to –15% on supply
Fast transient/burst
High energy pulse/transient
Damped oscillatory waves
Conducted disturbance, induced by radio frequency fields 150 kHz – 80 MHz
Electrostatic discharge 8 kV in air, 6 kV in contact
Power frequency magnetic fields 1000 Aeff/m
Damped oscillatory magnetic field 0.1 MHz and 1 MHz 30Ap/m
Radiated electro-magnetic field 80-1000 MHz, field strength 10 V/m
Harmonic current emitted into power source
Part 2 Industrial electro-magnetic field radiated in bands 0.15 – 30 /
30 – 1000 MHz
Special ultrasonic transducers with Ex certification are available. The
evaluation unit can also be delivered in a pressure-sealed housing. Please
see data sheets or ask your dealer.
UF_Multy8000_BA_02_1007
2 kV common, 1 kV normal mode
2 kV common, 1 kV normal mode
1 kV common, 0.5 kV normal mode