9 ROUTINE INSPECTIONS AND MAINTENANCE

9.1

DAILY

1. Check the detergent and neutralising agent levels in the cans and fill up if necessary.
2. Check spray arm rotation and inspect the water outlet holes visually to check that they are clean.
9.2

WEEKLY

1. Clean the sump filters, following the instructions provided.
2. Run PROGRAM 6, or another program with a thermal disinfection phase, without instruments for processing
in the appliance, for precautionary cleaning/disinfection of the washing chamber.
9.3

EVERY SIX MONTHS

1. Inspect the water solenoid valve intake filters; if necessary, clean them by pouring hot water through them in
the opposite direction to the operating flow.
2. Inspect the intake and outlet hoses of the detergent and neutralising agent pumps.
9.4

EVERY YEAR

Contact your nearest authorised Smeg service centre to have the appliance completely checked over at the end of the
warranty period and annually for successive years, or, if this occurs before the end of the year, when the "Appliance
maintenance" LED illuminates.
Maintenance operations are not covered by the product Warranty, which does not include the
replacement of components which deteriorate due to normal wear and tear.
The operations which Smeg authorised staff will carry out are:
1.
Inspection and if necessary replacement of worn peristaltic pump components (especially the internal hose)
2.
Inspection of the detergent intake pipes and replacement if necessary
3.
Inspection of the door gasket and replacement if necessary
4.
Inspection of the drying system filters (relative and absolute) and replacement if necessary
5.
Inspection and if necessary cleaning/replacement of the filters (water intake filters on the filler pipes, detergent
filters on the suction systems)
6.
Checking of the intake water hardness setting (the user must first have the intake water analysed to allow the
correct settings to be made)
7.
Checking of the detergent dosage settings.
8.
Inspection of the steam condenser assembly
a. Inspection of the nozzles, check that water is flowing correctly
b. Inspection of the connection, water intake, water drainage and level pressure switch connection
hoses
9.
Performance of a complete operating cycle, including the drying phase, to check for any leaks or malfunctions.
WARNING
SMEG declines all responsibility in the event of appliance malfunctions, or injury or damage,
arising from the failure to comply with the above recommendations.
User Manual
GW2145 - GW1160 - GW4060 - GW4190
Page 69 - 86

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Gw1160Gw4060Gw4190Gw01160s

Table of Contents