...3 ELECTRICAL INSTALLATION
3.3.2 Transmitters (All versions)
Caution. Unused
cable entries must
be blanked with
the
permanent
blanking
plugs
supplied with the

transmitters.

Caution.
Remove any exposed black conductive layer from the inner insulation of both coaxial
cables.
Substitute sensor cable of any kind is not acceptable.
Do not make connections except as shown.
Twist cable pairs together as shown.
Sleeve ALL bare wires.
Sensor cable may only be joined using company supplied junction box -
available separately.
Drive
Connections
Fig. 3.13 Sensor Cable Connections at the Transmitter (Remote version)
10
1
Slide
down
Pull out
slightly...
Fig. 3.12 Transmitter Connection Terminal access
Red
(CD 1)
Yellow
(CD 2)
Violet
Remove
Protection
Cover
4
2
2
...and
slide off
White Coax
Blue Inner (SIG 1)
Screen (DS 1)
Grey Coax
Pink Inner (SIG 2)
Screen (DS 2)
Ground Wire
(Safety Earth)
3
Slacken
captive
screws
Signal
Connections