3.4.6 Computer Connection – Fig. 3.22 and 3.23
Information. RS422/423 option is electrically isolated from all other MagMaster connections.
Fig. 3.23 RS 423 Connections (Single Ended or RS 232)
MagMaster
TERMINALS
TX-SIG
TX+SIG
RX-SIG
RX+SIG
Fig. 3.22 RS 422 Connections (Balanced)
MagMaster
TERMINALS
TX-SIG
TX+SIG
RX-SIG
RX+SIG
0VC
3 ELECTRICAL INSTALLATION...
RS422 Connection
NAME
RX DATA -
RX DATA +
TX DATA -
TX DATA +
0VC
SIGNAL GROUND
DCD
Link
DTR
together
DSR
MagMaster
Transmitter
TX - SIG
RX - SIG
TX + SIG
RX + SIG
0VC
RS232
9-Pin
Name
PC
Connector
RXD
2
TXD
3
GND
5
(linked)
(linked)
DTR
4
link
link
DSR
6
together
together
RTS
7
link
link
CTS
8
together
together
MagMaster
Transmitter
TX - SIG
RX - SIG
TX + SIG
RX + SIG
0VC
Link
APPLE Connector
(8 Pin MC)
5
8
3
6
4
7
Link
1
together
2
25-Pin
Hygienic
PC
Adaptor
Connector
Cable
3
Red
2
Blue
7
Yellow
(linked)
Green
20
link
6
together
4
link
5
together
15

Hide quick links: