...2 MECHANICAL INSTALLATION
2.3 Mechanical Installation
2.3.1 Transmitters
162mm (6.35in)
214mm
(8.43in)
32mm
(1.26in)
69mm (2.72in)
(Fixing Centres)
3 Fixing Holes,
6.5mm (1/4in) Diameter
Caution. Do not overtighten fixings,
especially on an uneven surface.
Fig. 2.16 Dimensions
2.3.2 Sensors
Caution
• Do NOT exceed the maximum working
pressure marked on the equipment.
• Use stainless steel (austenitic) bolts,
studs and nuts for flanged sensors below
200mm.
Fit gaskets
Gasket same size as pipe
Fig. 2.17 Gasket Fitting
4
Gasket
232mm (9.13in)
(Fixing Centres)
Allowance for
cable bends
200mm (8in)
Metal
face
PTFE
Gasket
<15mm bore sensors
Caution. For wafer type sensors of
<15mm bore, the fluid seal must be
made against the PTFE. Otherwise use
a full face gasket.
Fig 2.18 Wafer Type Sensors