Step 9 | Test reliability of the patterns
8 Ottobock | Myo Plus Quick Guide
Now test the reliability of the control in
real time in the menu
Each function in the prosthesis is assigned a colored outline, which can be
seen by scrolling down below the Spiderplot
By performing a movement with the phantom hand, the corresponding
muscle activities are shown in real-time through the gray filled shape
By checking or unchecking individual motions, patterns can be removed or
added to the Spiderplot to ease evaluation when necessary
If the user encounters problems during testing the recorded movements
try the following:
The user may have produced an incorrect pattern during or after the recording
Train the phantom movements with the user again, and try varying the
individual patterns
Simultaneously perform the movement with the contralateral hand
Take a break to regain the user's concentration and relax possible muscle fatigue
Record the patterns in the
By swiping to the left, the movement display will appear
The red bar indicates current relative contraction strength
The hand symbol indicates the movement pattern currently being detected.
A strong pattern is indicated by a solid
appearance . A weak pattern would flicker
between multiple movement pictures.
Show Spiderplot
Scroll down and select
Show Spiderplot instead of MotionClass
the view for a better evaluation of actual contraction pattern.
If the user can perform the movement patterns without major difficulty, you
can continue with the
basic set
Spiderplot / Training
test set
again
recording.
will change