Step 5 | Orientation of Myo Plus Cuff on limb
4 Ottobock | Myo Plus Quick Guide
Check connectivity of Myo Plus Cuff
After adjusting the Clinician settings navigate to the menu
Spiderplot/Training.
To test connectivity to electrodes and confirm all 8 are functioning, tap each
electrode set on the Myo Plus Cuff to confirm that patterns are created on
the Spiderplot.
The
Pattern Scaling Active
easier viewing.
Black electrode numbers indicate full functionality
White electrode numbers indicate a signal disturbances e.g. missing skin
contact or external influence. They will be turned off temporarily.
Red electrode numbers indicate a malfunction or broken electrode.
Pattern Scaling Active
Put on the Myo Plus Cuff so that the hand symbol is
legible from the user's point of view.
Then rotate the Myo Plus Cuff so that the ulna is positioned on
the indicated line between electrodes no. 1 and no. 8.
In this orientation, electrode numbers increase in the medial direction.
Electrodes 1-4 will be positioned over the extensors.
Electrodes 5-8 will be positioned over the flexors.
feature enlarges the size of the patterns for
Pattern Scaling In-Active