Step 7 | Prepare Test Set recording
Step 8 | Record Test Set
6 Ottobock | Myo Plus Quick Guide
Click on
Basic Set
in the main menu
Select pause time between recording sets; 4 seconds is default and
a good choice for most users.
For a test evaluation start with two recordings in the 90-degree bent
arm position – by clicking on the green arrow
is displayed.
Depending on which components are chosen, a number of
movement patterns must be performed during the
recording process:
Independently of the terminal device, three different movement
patterns must be performed in order: relax, hand open, hand close
With an active wrist rotation, two additional movement patterns must
be performed in order: palm up, palm down
1. Press
button to begin recording
Tip: the recording can be paused with
2. Relax Phase
Relax the limb during the
Complete relaxation is important so the system can recognize a baseline "quiet"
setting for the user.
TIP: Activate the Spiderplot in the recording circle, as shown below in step 5, to
verify that no muscle activity is occurring during relaxation.
3. Preparation Phase
Relax the Limb and prepare for the next indicated movement.
and continued at any time.
Blue
and
Red
phase.