Y-cable
10 Ottobock | Myo Plus Quick Guide
Connect both plugs of the Y-cable of the Myo cuff to the free
3-pin sockets of the extension cable. NOTE: Use silicone
grease (633F11) before connecting!s Cuff to the free 3-pin
socket of the extension cable.
extension cable
The cable with the blue dot belongs to the port
with a cable tie attached
Attention: Do not remove the cable tie!
2. Establish the connection
It is important to switch the cuff on before starting the Myo Plus app.
Do NOT turn on the demo dome yet.
Switch on the Myo Plus Cuff, you will hear 2 short
beeps and see a green light
Start the Myo Plus App and establish a Bluetooth connection
Login as a clinician
Go to menu
and MyoBock device and click on
NOW switch on the demonstration prosthesis within 15 seconds.
You will again hear 2 short beeps and see a green light.
If the Myo Plus Cuff is connected to the demonstration
prosthesis, the app switches to the main menu
and select
Now the terminal device will perform the user's movement
and select
Myo cuff
, choose between bebionic
Connect
Spiderplot / Training