2
3
4
3) Carefully pull the orthosis apart at the clos­
ure (see fig. 2).
4) Align the orthosis and place it onto the
baby's head (see fig. 3).
5) Check the fit of the orthosis:
→ The arrowhead on the front should
form a vertical line with the middle of
the nose.
→ The orthosis has to leave the ears and
eyes uncovered.
→ The lining should project beyond the
edge of the orthosis all around and
protect the skin against contact with
the plastic.
6) If the orthosis fits correctly: Hold the
orthosis carefully so it cannot shift and
fasten the closure. A gap in the closure is
normal and no cause for concern.
7) Face the flap of the elastic band up and
verify that the elastic band fits properly in
the guide of the closure.
8) CAUTION! To avoid injuries, always
open the closure before positioning the
orthosis.
If the orthosis does not fit correctly or
has shifted: Repeat the previous steps.