4.4 Information on Proximity to Certain Areas
CAUTION
Proximity to sources of strong magnetic or electrical interference (e.g. theft prevention
systems, metal detectors)
Injury due to unexpected behaviour of the prosthesis system caused by interference with internal
data communication.
Avoid remaining in the vicinity of visible or concealed theft prevention systems at the
entrance/exit of stores, metal detectors/body scanners for persons (e.g. in airports) or other
sources of strong magnetic and electrical interference (e.g. high-voltage lines, transmitters,
transformer stations, computer tomographs, magnetic resonance tomographs, etc.).
When walking through theft prevention systems, body scanners or metal detectors, watch
for unexpected behaviour of the prosthetic system.
CAUTION
Insufficient distance to HF communication devices (e.g. mobile phones, Bluetooth
devices, WiFi devices)
Injury due to unexpected behaviour of the product caused by interference with internal data
communication.
Therefore, keeping a minimum distance of 30 cm from HF communication devices is recom­
mended.
4.5 Information on Alignment/Adjustment
CAUTION
Operator errors during the adjustment process with the adjustment software
Injury due to unexpected product behaviour.
Participation in an Ottobock product training course is mandatory prior to using the product.
During this product training course you will receive a password giving you access to the
adjustment software. Additional product training courses may become necessary to qualify
for software updates.
Do not share your unlock PIN.
Use the online help which is integrated into the software.
CAUTION
Incorrect electrode settings
Injury due to unexpected product behaviour.
The electrodes are to be placed on intact skin only and with as much electrode-skin contact
as possible. In the case of strong interference from electronic devices, the position of the
electrodes should be checked and changed if necessary. If the interference cannot be elim­
inated or if you do not achieve the expected results by adjustment or selection of the appro­
priate control programme, please contact the Ottobock branch responsible for your country.
Set the electrode gain as low as possible in order to reduce interference from powerful elec­
tromagnetic radiation (e.g. visible or or concealed theft prevention systems at the
entrance/exit of stores), metal detectors/body scanners for persons (e.g. in airports) or other
sources of strong electromagnetic interference (e.g. high-voltage lines, transmitters, trans­
former stations, computer tomographs, magnetic resonance tomographs, etc.).