Chapter 1 Safety Warning and Instructions
Install an extraction system over the bath opening, if the operating of the equipment could
lead to liberation of hazardous air or gas mixture, for example the evaporation of the heat
transfer liquid! The rating of the fume hood shall be at least the maximum possible exhausting
temperature! If necessary for safety, additional temperature limiting and warning devices,
condensate accumulator shall be installed for the extraction system!
Caution: Note the following when connecting the equipment to an external application system:
1.
The external circulating pipe, including flexible tubing and hose, shall be connected only
after the equipment and application system are well seated and installed. The length and
distribution path shall be as short as possible and be so arranged that no kink and other
mechanical stress develops during normal use which otherwise may result in thermal and/or
liquid hazards!
2.
Flexible hose, fitting and coupling are important for the safe operation of the equipment;
Use only hose, fitting and coupling provided with the equipment or recommended by
Xtemp!
3.
The ratings of pressure and temperature for the hose, fitting and their resistance to the
chemicals shall meet the requirements of the operating temperature, pressure range of the
heat transfer liquid to be delivered for the jacket, container or heat exchanger of the
application system!
4.
Thermal insulation for hose, fitting, coupling and the application system are important for
performance and safety! The materials of the insulation shall meet the requirements of the
maximum operating temperature range!
5.
If no external circulating is necessary, use hose, fitting and coupling to short-circuit the liquid
connections as for an external application!
Keep the cables and wiring protected by conduit and/or hose and well secured; keep them
clearly away from direct contact with any exposed heated/cooled parts of the equipment
and application system!