Chapter 2 Installing and Operating
BJGW0141
Do not disassemble or exchange assembly parts and electric circuits by unauthorized personnel!
Hazard of electric shock!
In case of damaged shipment and dent, deformation or distortion of the enclosure, contact your
local distributor or supplier.

2.1.2 Installing

Install the equipment on a rigid, nonflammable surface and well-ventilated place. The bench
should be 600~800mm high for ease of operation. Keep enough space between wall and
back of the equipment, and keep clear of obstacles around the bottom of the equipment
for better ventilation. If the equipment is installed on an unstable surface or floor, abnormal
noise may be generated.
The voltage, frequency and current of the power supply must be rated at least the same as
those designated on the name plate of the equipment. The switch board is recommended
to be on the back right side of the equipment and equipped with outlet matching the
power plug of the equipment with protective ground (PE).
For possible wet location conditions, where cycle of evaporating and condensing, overflowing
or spillage may occur and for protection against hazard of possible electric shock, use
power socket with appropriate IP protection and additional residual current circuit breaker
(RCD) with rated breaking capacity!
Power socket, switchboard with circuit breaker or residual current circuit breaker for the
equipment shall be located closest to the equipment with good clearance for easy access,
so that immediate interruption is possible under emergent circumstance.
The minimum distance away from the ventilating holes at the back of the equipment to the
wall, furniture or other installations shall be 300mm! The inlets of fresh air and outlets of
exhausting air of the equipment shall be kept clear of any obstacles.
Keep the equipment away from hot-air emitting source, direct sunshine, strong magnetic field
and electric sparks!
Neither contamination of high concentration dust or corrosive gas, nor strong airflow surrounds!
Voltage, frequency and current for power supply must meet the requirements as designated
on the nameplate! Use power socket matching the power plug of the equipment. Hot line
(L)and neutral line(N)for single phase equipment must not be reversed!
XTEMP ENVIRONMENTS (SHANGHAI) CO., LTD.
http://www.xtemp-usa.com/ 
1
Bath cover assembly
/
31
DSGO0708-2-V211119/UXP8011-CD6400/CLASS I/NFL/RT+5~+100°C