1.3 Caution

Caution, Risk of burn hazard!
Do not touch or approach parts or area with high temperature, for example, bath cover and
the heat transfer liquid when maintained at operating temperature over 60°C!
Pull the temperature of the heat transfer liquid as close as possible to the ambient before
attempting to shut down the equipment, except when operations as such are prohibited for
the specimen! Warning, Hazards of burn and possible electric shock resultant from unattended
hot temperature or condensing and degraded dielectric strength!
The minimum distance away from the ventilating holes at the back of the equipment to the
wall, furniture or other installations shall be 300mm! The inlets of fresh air and outlets of
exhausting air of the equipment shall be kept clear of any obstacles.
Keep the equipment away from hot-air emitting source, direct sunshine, strong magnetic field
and electric sparks!
Neither contamination of high concentration dust or corrosive gas, nor strong airflow surrounds!
Voltage, frequency and current for power supply must meet the requirements as designated
on the nameplate! Use power socket matching the power plug of the equipment. Hot line(L)
and neutral line(N)for single phase equipment must not be reversed!
For possible wet location conditions, where cycle of evaporating and condensing, overflowing
or spillage may occur and for protection against hazard of possible electric shock, use power
socket with appropriate IP protection and additional residual current circuit breaker (RCD) with
rated breaking capacity!
Socket for power supply must be equipped with protective ground (PE) to prevent against
hazard of possible electric shock!
Connect the power from behind the equipment. Plug in and remove the cable by directly
holding the plug. No dragging of cable in any part. Protect the cable from being damaged by
contacting with the hot surface of the equipment or mouse etc!