Chapter 3 Controller Descriptions and Operation

3.9 Set default parameters

1) It may be used to recover data and functional parameters when system halts or any other
abnormal situation occurs.
2) Press
for more than 2 seconds until screen display Ht. Scroll the menu by
is displayed, then press
more than 2 seconds and release, once donE is displayed, it exits automatically.
Accidental power failure or interference, the abnormal situation will happen such as the
equipment may not be able to achieve normal temperature controlling, alarm may not be
cleared or abnormal temperature displays etc. Try
settings.

3.10 Terminate operating

1) Press
for more than 2 seconds. The circulating pump and the heater stop simultaneously,
and the screen displays oFF.
2) Disconnect the power by turning off the power switch, removing the power plug, or
disengaging the residual current circuit breaker (RCD).
Pull the temperature of the heat transfer liquid as close as possible to the ambient before
attempting to shut down the equipment, except when operations as such are prohibited for the
specimen! Warning, Hazards of burn and possible electric shock resultant from unattended hot
temperature or condensing and degraded dielectric strength!
Caution: Risk of freezing hazard!
When water is employed as heat transfer liquid, and when intended for standstill for long period
and when experiencing low ambient is possible, run the equipment at operating temperature
close to ambient for at least 1hr. Drain the bath containing water, and remove the water
completely. Disconnect the equipment and remove the mains plug from power source.
No draining of the heat transfer liquid with potential of chemical or biological hazard! Collect and
identify the heat transfer liquid for future use! Consult manufacturer or supplier of the liquid for
proper disposal!

3.11 Unit of temperature

1)
The default unit for temperature measurements and setting is degree centigrade.
2)
The unit of temperature may be changed to Fahrenheit or Centigrade as desired. Refer to
Table and item No. 4, use cF and similar steps for procedures.
XTEMP ENVIRONMENTS (SHANGHAI) CO., LTD.
http://www.xtemp-usa.com/ 
for less than 1 second ,the dEFt flashes. Press
/
31
 26
command first to recover default
dEFt
DSGO0708-2-V211119/UXP8011-CD6400/CLASS I/NFL/RT+5~+100°C
3
until dEFt
again for

Advertisement

Table of Contents
loading

This manual is also suitable for:

Uxp 8011-hc130Uxp 8011-hc250

Table of Contents