Constant Temperature Control - xtemp UXP 8011-HC090 Operating Instructions Manual

Immersion heating circulator
Table of Contents

Advertisement

 
Chapter 3 Controller Descriptions and Operation
get the desired value. Press
automatically in 4 seconds.

3.3 Constant temperature control

1) Connect the power by inserting the power plug, or engage the circuit breaker on the switch
board.
Turn on power switch and the controller runs self-check to display: 88888 , LED on, software
version number, AoFF ( Aon will show up if Auto is set to Aon ) settings for Ht , Lt . Refer to 3.5 for
procedures about Aon and AoFF setting. The software version number may change with
technique updates.
2) Check and/or set the working temperature, refer to item No.2 in the table.
3) If the equipment is in the standby state, press
pump will start to operate and the heater is on if the setting is higher than the actual
temperature.
4) When the temperature in the bath is lower than the set value, heating LED is on and the
equipment is in operation and start to heat. When the set temperature and actual
temperature are quite different, heating LED will be steadily lit with full power heating and the
temperature will go up with full speed. As the temperature approaches the set point, the LED
flashes as desired and the controller decreases the power for the heater. With the actual
temperature equals to the set temperature, the controller keeps the temperature by
constant power output.
5) To terminate the temperature control, press
pump and heater are stopped. The LEDs go off, and the display shows oFF .
Connect the power from behind the equipment. Plug in and remove the cable by directly
holding the plug. No dragging of cable in any part. Protect the cable from being damaged by
contacting with hot surface of the equipment or mouse etc!
Caution, Hazard of overflowing or low level!
Never attempt to initiate the equipment if no heat transfer liquid is filled or filling volume is lower
than the minimum level as specified; Never move the equipment if the bath tank is fully filled
with the heat transfer liquid; The liquid level may change with the heating and cooling of the
equipment, and with the immersion or removal of the specimen!
XTEMP ENVIRONMENTS (SHANGHAI) CO., LTD.
http://www.xtemp-usa.com/ 
to confirm and exit the SEt screen or exit
for more than 2 seconds. The circulating
/
31
DSGO0708-2-V211119/UXP8011-CD6400/CLASS I/NFL/RT+5~+100°C
 22
for more than 2 seconds, the circulating
3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Uxp 8011-hc130Uxp 8011-hc250

Table of Contents