Chapter 3 Controller Descriptions and Operation
recovered to higher than the Lt setting, the Lt alarm will disappear
automatically with seconds of delay, or may be cleared immediately by pressing
CooL: Control parameter for compressor, and select among three modes, Auto, on and oFF:
on:Compressor cooling with counter heating by PID, provided that the actual
temperature is below 40 °C.
Auto:Compressor on and off automatically. During heating up, once full heating is
necessary, the compressor will be turned off, until the actual temperature close to set
temperature. if the set temperature over 40 °C, and no cooling is required, the
compressor will also be turned off. If the set temperature is lower than the actual
temperature, and when the actual temperature below 40 °C, the compressor will be
switched on and keep working continuously.
oFF:The compressor is disabled, and the heater is regulated by PID.
Auto : Select between AoFF and Aon , which means manually starting and automatically starting.
Atc : Absolute temperature calibration (ATC), when the temperature displayed differs from the
temperature measured by a standard thermometer, ATC is used to calibrate the display.
AS : Timer for automatic startup. Valid only when the equipment is in standby mode. The
equipment will be initiated automatically once the set time elapses. The equipment may be
started immediately if
ASd : Timer for automatic shutdown. Valid only when equipment is in operating mode. The
equipment will be terminated automatically once the set time elapses. The equipment may
be terminated immediately if
cF: Shifting of the temperature measurement units. c means Centigrade and F means Fahrenheit.
cF setting affects all the parameters related to temperature, such as the display temperature,
setting temperature, Ht , Lt and so on. Relationship between c and F is as follows
°F=32+1.8×°C, °C=(°F-32)/1.8
For example: 0 °C=32 °F, 212 °F=100 °C.
dEFt :Return to the factory default setting. It may be used when system halts or any other
abnormal situation occurs.
ESc: Exit the SEt menu.
XTEMP ENVIRONMENTS (SHANGHAI) CO., LTD.
http://www.xtemp-usa.com/ 
is operated regardless of the remaining AS time.
is operated regardless of the remaining ASd time.
/
 18
31
DSGO0708-2-V211119/UXP8011-CD6400/CLASS I/NFL/RT+5~+100°C