Chapter 3 Controller Descriptions and Operation

Chapter 3 Controller Descriptions and Operation

3.1 Controller panel

3.1.1 Display for temperature and working parameters

Normally it displays the equipment temperature once switched on and initiated or oFF when
standstill.
Other possible symbols include: SEt , Ht , Lt , CooL , Auto , Atc AS , ASd , cF , dEFt , ESc , Stb .
Related definitions are as follows:
SEt : Setting of working temperature, must be 5 °C higher than low limit Lt and 5 °C lower than
the high limit Ht . The temperature setting beyond the scope will not be accepted.
Ht: This parameter is an absolute value and used for reminding only. When temperature in the
equipment is higher than the SEt value, Ht is displayed. The controller will trigger a beeping
alarm and disable the heating. While the actual temperature is higher than the Ht setting, the
Ht may be set to a higher value by pressing
actual temperature is recovered to lower than the Ht setting, the Ht alarm will disappear
automatically with seconds of delay, or may be cleared immediately by pressing
heating is possible only if the Ht is cleared.
Lt: This parameter is an absolute value and used for reminding only. When temperature in the
equipment is lower than the SEt value, Lt is displayed. The controller will trigger a beeping
alarm. While the actual temperature is lower than the Lt setting, the Lt may be set to a
lower value by pressing
XTEMP ENVIRONMENTS (SHANGHAI) CO., LTD.
http://www.xtemp-usa.com/ 
Fig 6 Controller panel
60 Reset button (for resetting when alarming, the
circle indicating the location)
61 Display for temp and working parameters
62 Status indicators
63 Key board
64 Power switch
so that Lt beeping can be cleared. If the actual temperature is
/
31
 17
so that Ht beeping can be cleared. If the
DSGO0708-2-V211119/UXP8011-CD6400/CLASS I/NFL/RT+5~+100°C
3
. The