OPTO-EDU (BEIJING) CO., LTD. Content Overview ....................4 1.1. Product Features ..................4 1.2. Main uses and scope of application ............5 1.3. Model & Product Name .................5 1.4. Composition of the model No. and its representative meaning ..5 1.5. Main technical data ................6 1.6.
Page 3
OPTO-EDU (BEIJING) CO., LTD. 2.4. Vacuum system ..................13 2.5. Electrical cabinet control ..............14 2.5.1. Electrical cabinet ................14 2.5.2. Computer System................15 Use and operation of electron microscope software ..... 17 3.1. Familiar with the basic use of Windows 7 ........17 3.2.
Page 4
OPTO-EDU (BEIJING) CO., LTD. 4.3. Monitoring and Recording in Operation ..........38 4.3.1. Working of the vacuum system ...........38 4.3.2. Software Working Status ..............39 4.4. Operating procedures, methods & precautions for shutdown ..39 4.4.1. Daily standby status ..............39 4.4.2. Shutdown Status ................39 4.4.3.
OPTO-EDU (BEIJING) CO., LTD. Overview 1.1. Product Features A63.7080 Scanning Electron Microscope Std FEG SEM, is a digital Field Emission Gun scanning electron microscope (referred to as A63.7080). It is a large- scale precision electronic optical instrument for studying the microstructure of objects.
OPTO-EDU (BEIJING) CO., LTD. formats, and conversion between different image formats. The optional image processing software enhanced a series of image processing functions such as particle size analysis, on the basis of standard image processing software. 1.2. Main uses and scope of application...
OPTO-EDU (BEIJING) CO., LTD. 1.5. Main technical data Resolution 1.5nm@ 15kV or 30kV Magnification 15x~500000x Electron Gun Schottky Field Emission Gun Accelerating Voltage 0~30kV Lens System Electromagnetic lens Objective Diaphragm Include 3 diaphragm, out of vacuum room, position customized Specimen Room Include 6 interface, can connect WDS、 EDS、 BSE and etc.
OPTO-EDU (BEIJING) CO., LTD. 1.7.3. Requirements for power supply 1.7.3.1. Requirement for power A63.7080’s main body (excluding X-ray analysis equipment), need to supply: AC 220V, 50Hz, 15A, with UPS power supply, at least two power output sockets (220V, 50Hz, 10A and 220V, 50Hz, 16A respectively) ). The experimental indoor air equipment switch is: AC 220V, 40A.
OPTO-EDU (BEIJING) CO., LTD. equipment is in your environment (adjacent, upstairs, downstairs, etc.), it is good for measuring or correcting problems. 1.8. Impact on the environment and energy The X-ray radiation of A63.7080 meets the standards required by the state and does not affect the environment and energy 1.9.
OPTO-EDU (BEIJING) CO., LTD. System composition and working principle 2.1. Overall structure A63.7080 interacts with an electron beam with a certain energy to excite a physical signal. The imaging signal is collected, digitized, and the like by means of a computer system and an electrical control system to form an image.
OPTO-EDU (BEIJING) CO., LTD. 2.2.1. Schottky field emission electron gun The Schottky field emission electron gun is a high-brightness electron source consisting of a very fine single crystal tungsten wire tip and four electrodes. The four electrodes are the suppression pole, the absorption pole, the focusing pole and the anode, and the tungsten wire tip acts on the electric field.
OPTO-EDU (BEIJING) CO., LTD. 2.2.6. Signal Detecting System The signal detection system collects low energy secondary electrons and high energy backscattered electrons. When the grid at the front end of the detector is applied with a positive voltage, low-energy secondary electrons are attracted into the detector, and a positive voltage of 10KV is applied to the scintillator to accelerate the secondary electrons.
OPTO-EDU (BEIJING) CO., LTD. standard sample stage has a similar sample stage operation and the range of motion is as follows:: Horizontal X,Y=50mm Vertical Z=25mm Inclination T=-5°~ +90° Rotation R=360°Continous The sample locking device can be used to lock the sample table and the sample chamber after the sample stage is inserted to reduce the effects of vibration.
OPTO-EDU (BEIJING) CO., LTD. the sample under different conditions. 2.4. Vacuum system The electron optical lens barrel and the sample chamber must be operated under a certain vacuum state, and the realization and control of the entire system vacuum is completed by the vacuum system. The vacuum system is mounted below the sample chamber and is connected directly to the barrel via a vacuum line.
OPTO-EDU (BEIJING) CO., LTD. Field Emission Gun Ion Pump 1 Transition Chamber Ion Pump 2 Condenser Valve 1 Electron Optical Tube Objective Sample Chamber Valve 3 Vacuum Meter Vacuum Meter Vibration Reduction Molecular Pump Mechanical Pump Figure 2-2 Vacuum System The vacuum control part is installed on the front and rear panels of the main frame, and can be fully controlled from low vacuum to high vacuum.
OPTO-EDU (BEIJING) CO., LTD. parts: SEM control electric system, vacuum power supply and field emission gun power supply. The power switch of the SEM control electrical system, the power switch of the vacuum system and the power switch of the field emission gun are relatively independent and are controlled by respective power button switches.
Page 17
OPTO-EDU (BEIJING) CO., LTD. and image format selectable settings; supports a variety of common image formats, and conversion between different image formats; the software integrates some common image processing functions, which is more convenient to use. The computer operating system can be installed with electronic mirror image processing software, which is a more powerful independent image processing software;...
OPTO-EDU (BEIJING) CO., LTD. Use and operation of electron microscope software 3.1. Familiar with the basic use of Windows 7 The A63.7080 operating software is installed on the Windows 7 operating system and complies with all applications of the Windows 7 operating system.
OPTO-EDU (BEIJING) CO., LTD. 3.2.1. Software Scan Control Window Figure 3-3、3-4 show main function & meaning of software Menu bar Key bar Function field Main field Status field Figure 3-3 Scan Control Window www.cnoec.com.cn Page 18 of 51 sale@cnoec.com...
Page 20
OPTO-EDU (BEIJING) CO., LTD. Figure 3-4 Scan Control Function Note Mouse operation: After clicking the active block on the toolbar, drag the mouse to fine tune the corresponding parameters; After double-clicking the active block on the toolbar, drag the mouse to thicken the corresponding parameters;...
OPTO-EDU (BEIJING) CO., LTD. 3.2.2. Starting and Shut down Figure 3-5 Click“SE” 1. Run the SEM operating software. 2. Click “SE” in the upper right corner of the control interface to open the SEM operating software. (The configuration information when the last use is opened by default) Figure 3-6 Click“VAC”To Closc...
Page 22
OPTO-EDU (BEIJING) CO., LTD. Table 3-1 Function Keys Meaning www.cnoec.com.cn Page 21 of 51 sale@cnoec.com...
OPTO-EDU (BEIJING) CO., LTD. 表 3-1(续) 3.2.4. Electronic Gun Control The main function area of the control software has multiple tabs (Beam, Align, Image, AUX, Status), and the Beam tab is the gun control area, as shown in Figure 3-8. Divided into vacuum control zone, gun parameter setting zone, error prompt zone, including all functions of electronic gun parameter setting and monitoring.
Page 24
OPTO-EDU (BEIJING) CO., LTD. Vacuum Settin Gun Setting Error Hint Figure 3-8 Electron Gun Function Window Figure 3-9 Advance Function Window www.cnoec.com.cn Page 23 of 51 sale@cnoec.com...
OPTO-EDU (BEIJING) CO., LTD. The advanced function window includes four pages: Power Set, Power Monitor, Diagram, and Options. On the Power Set page, you can fine-tune the gun parameters (recommended for users). The Power Monitor page displays the current filament current value and the voltage values of the electrodes. The Diagram page displays the current vacuum valve status.
OPTO-EDU (BEIJING) CO., LTD. 表 3-2 电对中功能 3.2.6. Image electrical displacement Figure 3-4 电位移调节 3.2.7. Main function area The main function area of the scan control window is shown in Figure 3 12. In the function area, the function bar for adjusting brightness, contrast, adjusting focus and magnification is included, which can easily adjust various parameters.
Page 27
OPTO-EDU (BEIJING) CO., LTD. Figure 3-5 Scan Control Window Table 3-3 Scan Control Function www.cnoec.com.cn Page 26 of 51 sale@cnoec.com...
OPTO-EDU (BEIJING) CO., LTD. 3.3. Image acquisition 3.3.1. Overview of the Software Interface The software control interface is shown in Figure 3 13, including the menu bar, toolbar, Figure image display area, and scan control area. After opening the control software, follow the corresponding function adjustment to get the Figure image in the figure image display area.
OPTO-EDU (BEIJING) CO., LTD. Figure 3-6 Software Control Window 3.3.2. Scan Settings During the adjustment of the Figure image, right-click on the Figure image area to see the interface shown in Figure 3 14. Click on the corresponding option to make the relevant settings. Click on "settings" to set the detector type, scan speed, sync mode, etc., as shown in Figure 3 15.
OPTO-EDU (BEIJING) CO., LTD. 3.3.3. File Management Menu Click on the menu bar "file" to save or print the Figure image as shown in Figure 3 16. There is a shortcut to save the Figure image in the toolbar. As shown in Figure 3 17, you can select the storage location and name the Figure image file.
OPTO-EDU (BEIJING) CO., LTD. 3.3.4. Getting images The toolbar of the software interface has a shortcut to scan the Figure image, as shown in Figure 3 19, and the functions are shown in the table. Figure 3-12 Scan Button Table 3-4 Scan Image Button Meaning www.cnoec.com.cn...
OPTO-EDU (BEIJING) CO., LTD. SEM Microscope Operation Procedure 4.1. Preparation and inspection before use 4.1.1. Power Supply Check Check whether the power supply (AC 220V) is normal, the cable connection is correct, and the ground wire is connected properly before A63.7080 is turned on.
OPTO-EDU (BEIJING) CO., LTD. electron microscope. Just open the V1 block valve, the electron microscope control software will automatically adjust the brightness, contrast, auto focus and other operations. You can immediately see the figure image in the figure window, which is easy to operate. Fast. To start the electron microscope from the off state, you need to click the "hot start"...
OPTO-EDU (BEIJING) CO., LTD. 4.2.2.3. Handling precautions In order to prevent the instrument from malfunctioning, the instrument is equipped with a chain protection device. However, some incorrect operation will still cause the instrument to malfunction. To this end, in addition to strict...
OPTO-EDU (BEIJING) CO., LTD. shutoff valve V1. 2. Click “Deflating” on the software interface and the vacuum system will automatically deflate the sample chamber. 3. After the deflation is completed, the sample chamber door is loose and the sample chamber is in an atmospheric state.
OPTO-EDU (BEIJING) CO., LTD. 4.2.5. How to get images After starting the electron microscope, open the V1 block valve and you can see the figure image in the figure window of the software control interface. In actual use, in order to obtain the desired Figure image, relevant adjustments, such as axis, astigmatism, brightness contrast, etc., are required.
Page 37
OPTO-EDU (BEIJING) CO., LTD. When adjusting the objective lens to close the axis, follow the steps below: 1. Use the active selection and fast scan mode to create a “feature point” on the Figure image. 2. Select a range of about 5000 times the low magnification; click the wobbler button to observe the movement of the image.
OPTO-EDU (BEIJING) CO., LTD. 4.2.6.4. Sample tilt T Rotate the knob on the door of the sample chamber marked with “T”, the sample is tilted, the angle of inclination is 0-90°, and there is a scale indication. 4.2.6.5. ample Grounding The sample forms a loop through the sounder and ground.
OPTO-EDU (BEIJING) CO., LTD. machine should be stopped according to the operation steps, and the manufacturer should be notified in time for maintenance. 4.3.2. Software Working Status For details, see the software log. Operating procedures, methods & precautions for shutdown 4.4.
OPTO-EDU (BEIJING) CO., LTD. Maintenance A63.7080 requires very limited user maintenance and is very demanding. Under normal circumstances, the user cannot open the panel to prevent the high voltage and the circuit board from being exposed to the outside, causing damage to the operator. At the same time, if there is no guidance from the manufacturer's engineers, the user cannot move the overall position of A63.7080 to avoid damage to the instrument and make...
OPTO-EDU (BEIJING) CO., LTD. 7. Install the sample stage on the sample chamber, push in gently, and click “Pump”. 8. 8. Wait until the vacuum system has completed the ultra-high vacuum and other work can be performed. 5.1.2. Cleaning the objective diaphragm and rod After working for a period of time A63.7080, the objective diaphragm is...
Page 43
OPTO-EDU (BEIJING) CO., LTD. blower and place it on a clean, lint-free cloth. 7. Use a bamboo stick to remove the rubber “O” ring from the light rod. 8. Apply a small amount of polishing paste on a lint-free cloth, polish the polished rod, especially in the groove where the light is placed, and carefully polish it with a thin bamboo stick against the lint-free cloth;...
Page 44
OPTO-EDU (BEIJING) CO., LTD. redness, and the dirt on the visible light ray changes from yellow to black, and then gradually disappears. Repeat the burning for 1-2 times, each time for a few seconds, until the surface of the enamel is smooth, be careful not to burn for a long time, so as not to damage the glare.
OPTO-EDU (BEIJING) CO., LTD. 5.1.3. Other maintenance work Maintenance of A63.7080 also includes cleaning the liner, cleaning the scanning coil, the objective lens lower boots and the entire electron optical tube, or moving the A63.7080 to other places. These maintenance work must be performed by a professional engineer appointed by the manufacturer.
OPTO-EDU (BEIJING) CO., LTD. Transportation and Storage 6.1. Lifting and transportation precautions A63.7080 should prevent inversion, rain and strong earthquakes during hoisting, handling and transportation to prevent severe cold and exposure. 6.2. Storage conditions, storage period and precautions A63.7080 should be stored in a dry, ventilated, non-corrosive gas house before opening the box to prevent inversion, rain, cold and exposure.
OPTO-EDU (BEIJING) CO., LTD. Unpacking and Inspection 7.1. Unpacking Precautions 1. Before opening the A63.7080, carefully check the laboratory conditions to ensure that the installation and commissioning requirements are met. 2. The A63.7080 unpacking environment is guaranteed to be dry, ventilated and free of corrosive gases.
OPTO-EDU (BEIJING) CO., LTD. 7.2.2. Instrument appearance inspection Under the condition of sufficient light, visual observation, the appearance of the instrument should be free of large scratches and stains. www.cnoec.com.cn Page 47 of 51 sale@cnoec.com...
Address : F-1501 Wanda Plaza, No. 18 Shijingshan Road, Beijing 100043, China 8.2. User feedback In order to improve product quality, OPTO-EDU conscientiously implements the ISO9000 quality system improvement requirements. After the installation, acceptance and training of the instrument, please fill out the “Customer Satisfaction Information Questionnaire”...
OPTO-EDU (BEIJING) CO., LTD. Appendix: Abbreviations, Glossary of Terminology Abbreviations & Glossary of Terminology Chinese English 解释 Refers To 扫描电镜 Scanning Electron Microscope 场发射 Field Emission Gun A63.7080 A63.7080 A63.7080 Scanning Electron Microscope 镜筒 COLUMN Electron optical tube column 分辨率...