Page 1
A14.1063, A14.1064 Inverted Biological Microscope A16.1063, A16.1064 Inverted LED Fluorescent Microscope Instruction Manual To ensure the safety and obtain satisfactory performance, please study this instruction manual thoroughly before start to use your instrument.
Before Using Safety Symbol Symbol Description The surface is hot and cannot be touched with bare hands. Please read this manual carefully before using the product. Incorrect usage can cause personal injury or property damage. The main power switch is turned on. The main power switch is turned off.
5. Grounding the machine to avoid lightning strike. To ensure safety, always make sure the main switch is in "O" (disconnecting) and cut off the power while waiting for the LED lights and the lamp room to cool completely. Input voltage check: the input voltage indicated on the back of the microscope is consistent with the supply voltage, otherwise the microscope will be seriously damaged.
Page 6
Because the solvent such as ether and alcohol are extremely flammable, do not operate power switch of various electrical equipment when using, and keep away from open fire, please ensure indoor ventilation. Do not use organic solvents to clean the non-optical parts of the microscope.
3.2 Turning the Illumination on switch ( for 3.2.1 Manual lighting Illumination A14.1063 series) switch Make sure the power indicator is on and press the light switch. 3.2.2 ECO switch (Optional functions) As shown in figure, the induction device can sense objects within 0.5m in front of...
3.3 Adjusting the Illumination Brightness As shown in the direction of rotation, the light source is enhanced, whereas the light source weakened. 3.4 Specimen Placement and Stage Knob Adjustment Load the slide holder into a mechanical stage, cover glass slides downward slowly into the grooves of the slide plate.
3.4.2 Holders There are a variety of holders to choose from. Model Terasaki micro plate Universal holder 54mm Petri dish holder/ Slide holder 60 mm Petri dish holder/ Slide holder 90 mm Petri dish holder 3.4.3 Auxiliary stage Connect the auxiliary stage to the left and Auxiliary stage right of the stage to expand the stage width.
3.6 Focusing on the Specimens limit screw 3.6.1 Focus knob To adjust the focus, rotate the focus knobs on the right and left sides of the microscope to move the objective up and down. The figure to the right illustrates the relationship between the rotational direction of the focus knobs and the vertical notion of the objective.
3.7 Adjusting the Transmitted Illumination Aperture Diaphragm Insert a centering telescope in the eyepiece tube, adjust the focal length, and view the objective pupil plane (a bright circle) and the aperture diaphragm image. Adjust the aperture diaphragm so that the size of the Aperture diaphragm diaphragm image is 70-80% of the size of open/ close lever...
3.9 Operation the Phase Contrast Microscope Phase contrast microscopy requires phase contrast (PH) objectives and PH sliders. Annular diaphragm Hollow position Centering hold Centering hold (on the side) (on the side) If PH Slider is used, it is possible to center the annular diaphragm. Turn the centering screw of the PH slider to merge the annular diaphragm image with the phase ring image in the objective.
3.10 Operation the Emboss Contrast Microscope To perform emboss contrast microscopy, a condenser-side emboss contrast slider and eyepiece-tube-side emboss contrast slider must be mounted on the microscope. 3.10.1 Condenser-side emboss contrast slider The condenser-side emboss contrast sector slider equipped with sector diaphragm diaphragm.
3.11 Operating the Episcopic Illumination Microscopy (for the NIB410-FL Only) ( 3.11.1 Episcopic illumination A14.1063-FL) As right figure shown, press the illumination switch switch between episcopic illumination LED and diascopic illumination fluorescent LED mode. Fluorescent illumination will turn weaker, when...
position, seen from front microscope: If you select the bright-field position, the filter cube is removed from the optical path. The three bright-field positions have the same function. SIGN Rotate the turret so that the fluorescence positions indicated by the arrows on the left and right are the same (all points to “1”...
3.12 Operating the LCD(for A14.1064 series) 3.12.1 The status of LCD Start & working mode Standby SLEEP Lock Time setting TIME 3.12.2 Operating the illumination switch Click brightness adjuster:becoming standby mode,it’s“SLEEP”in the screen。Click again,it becomes working mode。 Press brightness adjuster three seconds:setting time. First minute scale figures flash, then click the button, hour scale figures become flashing.
Chapter 4 Assembly 4.1 Installation diagram The order of installation of each component is shown in the following figure, and the digital indicate the installation step. Save the supplied hexagonal spanner. You need it when you change parts. CAUTION Before installation, make sure all the parts are free of dust and dirt. Do not scratch any part or surface of the glass.
4.2 Installation Steps 4.2.1 Installing the Condenser Unscrew the captive screws on the condenser certain distance before aligning the tube with the dovetail groove of the condenser, then pushing lightly to the lowest position horizontally. and tighten the set screws. Set screw 4.2.2 Installing the Phase Contrast Slider With the sign facing up, push the phase...
4.2.4 Installing the Objective 1. Adjust the coarse focus knob until the converter drops to low limit. 2. Screw the objectives with the lowest magnification to the converter from the left or right and then push the nosepiece clockwise to install the other objectives in descending order of magnification.
4.2.7 Installing the Camera Remove dust cover from 1X relay 1X Relay lens lens. Screw the camera into the relay lens Dust as shown. cover In the process of installing the camera, pay attention to one hand always hold the camera to prevent fall and break.
4.3 Assembly for Episcopic Illumination Microscopy 4.3.1 Attaching the Filter Cube Remove the cover from the filter cube mounting port. Turn the turret to a position where a Filter cube filter cube can be mounted easily. Check the address on the turret, and align the filter cube with the guide Cover groove before inserting.
Chapter 5 Maintenance and Storage 5.1 Cleaning Clean or disinfect the lenses and other components according to the following instructions. Cleaning tools Blower Soft brush Soft cotton cloth, lens cleaning tissue, gauze, etc. Absolute alcohol (ethyl alcohol or methyl alcohol), medical alcohol ...
Cleaning off Immersion Oil 5.1.3 (1) Wipe off using petroleum benzene. (2) Then, wipe off using absolute alcohol (ethyl alcohol or methyl alcohol) for a better finish. If petroleum benzene is unavailable If petroleum benzene is unavailable, use methyl alcohol. Note that methyl alcohol is less effective, and requires more wipes.
Chapter 6 Troubleshooting 6.1 Optical System TROUBLE CAUSE SOLUTION The nosepiece is not in the Locate the nosepiece properly located position (objective and The edge of the where it clicks light path not coaxial) field of view is dark or the brightness is not uniform A lens (the objective, condenser, Clean it thoroughly...
6.2 Mechanical System TROUBLE CAUSE SOLUTION image cannot focus Specimen placed Turn it over; with the high-power reversely; Use standard cover glass with objective Cover slip is too thick thickness of 0.17mm The objective will Specimen placed Turn it over; touch the Specimen reversely;...