Procedure
1. Isolate the device from power.
2. Remove the housing cover. To do so, unscrew the two nuts on the rear panel of the device and
push the housing cover to the back.
3. If applicable, remove the option module blank plate from the rear panel.
4. Loosen the two screws
5. Insert the option module
6. Fasten the option module
Tighten the screws with a torque of 1.3 Nm.
7. Insert the printed circuit board holder
Tighten the screws with a torque of 2.5 Nm.
7.8
Wiring option modules
Overview
①
②
Figure 7-10
Note
ULTRAMAT 7: Deviating position of the OM2.x connections
With ULTRAMAT 7 analyzer modules, the connector of the option module is not on the printed
circuit board but is on an additional circuit board. This is mounted horizontally at the upper end
of the analyzer module.
Rack-mounted device
Operating Instructions, A5E31874006-12
Installing/removing and connecting analyzer and option modules
③
Analyzer module printed circuit board
Interface
Connecting option modules based on the example of an OXYMAT 7
and remove the printed circuit board holder
①
at the appropriate mounting location into the device.
①
④
with both screws
②
again.
7.8 Wiring option modules
②
to the rear panel.
.
65