13.3
Technical specifications of analyzer modules
13.3.1
OXYMAT 7 technical specifications
Table 13-4
OXYMAT 7: General technical specifications
General
Information on the smallest span
Max. power consumption
Weight (only module)
Electrical safety
Max. EMC error during operation with one and/or two
OXYMAT 7 / ULTRAMAT 7 units
The high-temperature version of the analyzer module is not intended for installation in the rack-mounted device.
1)
Table 13-5
OXYMAT 7: Measuring ranges
Measuring ranges
Number of measuring ranges
Parameters can be as‐
signed in the measur‐
ing ranges
The high-temperature version of the analyzer module is not intended for installation in the rack-mounted device.
1)
Note
Temperature change for high-temperature OXYMAT 7
If you change the temperature range, the device has to be re-calibrated. Detailed information
can be found in the operating manuals, see "Bibliography".
Table 13-6
OXYMAT 7: Gas inlet conditions
Gas inlet conditions
Sample gas pressure
Standard devices with hoses
Tolerance within the valid pressure range
Standard devices with hoses with external
reference gas pump
Tolerance within the valid pressure range
Standard devices with pipes
Rack-mounted device
Operating Instructions, A5E31874006-12
Warm-up phase
Smallest possible measuring
span
Largest possible span
500 to 1 500 hPa (absolute)
± 150 hPa
700 to 1 200 hPa (absolute)
± 150 hPa
500 to 3 000 hPa (absolute); short-term < 5000 hPa (abs.)
13.3 Technical specifications of analyzer modules
OXYMAT 7
See module nameplate or device certificate
≤ 60 W
< 5.5 kg
In accordance with DIN EN 61010-1
OXYMAT 7
Max. 4; parameters can be as‐
signed freely
0.5% O
2
100% O
2
Technical specifications
OXYMAT 7 high-tempera‐
ture
1)
≤ 120 W
< 7.5 kg
1.0% according to EN
61326-1 (2013)
OXYMAT 7 high-temperature
Max. 4; parameters can be as‐
signed freely
2% O
2
100% O
2
1)
119