Combined Operation; Slots For Analyzer Modules; Operating Modes; Setting Note - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Combined operation

5.1

Slots for analyzer modules

The following figure shows the rear of the rack-mounted device with two slots for the analyzer
modules.
Figure 5-1
5.2

Operating modes

5.2.1

Setting note

Note
Operating mode-specific parameter assignment of the gas path in serial/parallel operation
Analyzer modules can be operated in series or in parallel.
• If you are only using one analyzer module or want to operate two analyzer modules in series,
set the value "Shared" in the "Gas path" parameter field.
• If you are operating two analyzer modules in parallel, however, select the setting "Separate".
You can find the corresponding setting options in the menu → [2.12].
Rack-mounted device
Operating Instructions, A5E31874006-12
Slot 1 (AM1), example OXYMAT 7
Slot 2 (AM2), example CALOMAT 7
Slots for analyzer modules, rack-mounted device rear
5
33

Advertisement

Table of Contents
loading

Table of Contents