3.2.4.3.15 If the main relative permissible
error limit of DER measurement does not meet
the requirements of 3.2.4.3.14 of this OM, the
dosimeter cannot be verified and should be
sent for repair.
3.2.4.4 Determining the main relative per-
missible error limits of photon ionizing radia-
tion DE measurement
3.2.4.4.1 Follow the steps in 3.2.4.3.2 and
3.2.4.3.3 of this OM. Fix the dosimeter that