ionizing radiation DER at a 30-second inter-
val. Calculate the main relative permissible er-
ror limit of measurement results as covered in
3.2.4.3.7 – 3.2.4.3.10 of this OM.
3.2.4.3.12 Carry out information exchange
with the dosimeter using the "Read" procedure
available in technological software. Make sure
that the dosimeter has switched from
the "Operation" to the "Standby" mode.
131