13. Proprietary Communication Protocol
Communication Protocol (TRCN-443):
1) Proprietary:
Order:
ESC
ID
P
CR
LF
0x1B
0x50
0x0D
0x0A
Hexadecimal
The ID is configured at the instrument from 1 to 32.
Answer:
L > > V V V V V V N T U C C C C C C m A
Example of answer for an un-stable value (for low turbidity 0 – 10NTU)
L > 0.001 N T U 1 2 . 0 0 m A
Example of answer for a stable value (for low turbidity 0 – 10NTU)
L > > 0.001 N T U 1 2 . 0 0 m A
Note: a) When the answer comes a C instead of a L, it means that the equipment is under
calibration function (is being operated in location by the user, executing the calibration
operation at the instrument).
b) When the answer comes an S instead of a L, it means that the equipment is under Set Up
function and it is being operated in location by the user.
2) RS485 – It is a "physical location", where the proprietary protocol will be "transported".
As factory default, this communication comes configured as:
Speed = 9600
Parity = none
Number of Bits = 8
Stop Bit = 1
ID = 1