RCP TC-2 and RCP TC-3 LED status indication
and troubleshooting
LED
Green
Flashing green
Amber
Flashing amber
Flashing red
WARNING: The RCP TC-2 or RCP TC-3 plate may be hot.
Issued 06 2022
Status
Power on, probe present, temperature within
target
Power on, probe present, temperature
outside of target
Power on, probe not present, temperature
within target
Power on, probe not present, temperature
outside of target
Error
Troubleshooting
n/a
Wait for the rack to reach stable temperature.
If the probe is present, clean the port and
probe contacts.
If the probe is present, clean the port and
probe contacts. Wait for the rack to reach
stable temperature.
Cycle power to restart. If the problem
persists, contact your supplier.