Installation
!
18
3.4.5
General instructions
Normally use rod probes. Rope probes are used in liquids for measuring ranges > 4m and with
restricted ceiling clearance which does not allow the installation of rigid probes.
Coax probes are not influenced by the installation conditions. They may also be operated
• in the filling curtain
• in arbitrary proximity to internal fittings
• at viscositys up to 500 cSt.
Mounting location
• Do not mount rod or rope probes in the filling
curtain (2).
• Mount rod and rope probes away from the
wall (B) at such a distance that, in the event of
build-up on the wall, there is still a minimum
distance of 100 mm between the probe and
the build-up.
• Mount rod and rope probes as far away as
possible from installed fittings. "Mapping "
must be carried out during commissioning in
the event of distances < 300 mm.
• Minimum distance of probe end to the
container floor (C):
– Rope probe: 150 mm
– Rod probe: 100 mm
– Coax probe: 10 mm
• When installing outdoors, it is recommended
that you use a protective cover (1) see
Accessories on → ä 65.
Note!
Seal for devices with G 1½" thread
The thread and the type of seal comply to DIN 3852 part 1, screwed end type A. Gaskets according
to DIN 7603 with a dimesnion of 48x55mm can be used for it. Please use a gasket according to this
standard in the form A, C or D and of a material that is resistant to your application.
Levelflex M FMP45 with HART/4...20 mA
1
B
C
Endress + Hauser
2
L00-FMP4xxxx-17-00-00-xx-007