Erasing Data; Erasing Single Data Points - Hach DR/2010 Instrument Manual

Hide thumbs Also See for DR/2010:
Table of Contents

Advertisement

Figure 17
Example Of A User Program Printout
3.8

Erasing Data

3.8.1 Erasing Single Data Points

DR/2010 User Program Report
Printed 26-MAR-96 11:32:55
Adj Chloride Test - Program Number: 968 Last Calibrated: 23-JAN-96 09:21:00
Default Measurement Form:
Alternate Form 2: 0.000 mg/L NaCl
Alternate Form 3: 0.000 mg/L EDL
High Limit: 8.97 mg/L Cl
Low Limit: 0.05 mg/L Cl
Monochromator Wavelength: 420 nm
Standard
Concentration
1
0.05 mg/L Cl
2
0.20 mg/L Cl
3
0.50 mg/L Cl
4
1.00 mg/L Cl
5
2.00 mg/L Cl
6
5.00 mg/L Cl
7
10.00 mg/L Cl
Calibration Curve Forced Through Zero? Yes
Calibration Style: C = a + bA + cA
a =
0.000
b =
0.1050
c =
0.0052
d = - 0.0002
2
r
=
0.9994
Timer 1: 3:30
Timer 2: 5:00
You may only erase stored data after it is recalled using the
recall data, see 3.6, Recalling Data. Erasing stored data may take up to 35
seconds. The instrument will not display an "in progress" message. Please
leave the instrument on during the process.
1. When the desired group and sample number are displayed, press
. The display will show (i.e., G100 is the group number
ERASE
assigned; #3 is the sample number):
&&&&
$
2. Press
. The display will show
ENTER
displayed group and sample number.
SN:789456123
00.00 mg/L Cl
Absorbance
2
3
+ dA
#"
&&&
-H=IA/!
ERASED
57
Factor: 0.1420
Factor: 1.000
0.007
0.028
0.070
0.140
0.281
0.700
1.400
RECALL
nm
and then return to the
key. To
SHIFT