HP 8924C Assembly page 604

Mobile station test set
Hide thumbs Also See for 8924C:
Table of Contents

Advertisement

Chapter 14, Diagnostics Descriptions
Description Of Miscellaneous Diagnostics (MSDIAGS3)
RF Input/Output Test
Table 130
Measurement
The RF Generator is set to 100 MHz at −100 dBm to
Duplex Out to
RF In/Out
the DUPLEX OUT. The user must connect a cable
from the DUPLEX OUT to the RF IN/OUT connec-
tor. The power meter is zeroed. Then the RF genera-
tor's amplitude is set to +10 dBm and the power meter
is read.
Duplex Out to
The RF generator is set to 100 MHz at +10 dBm to the
Ant In
DUPLEX OUT. The user must connect a cable from
the DUPLEX OUT to the ANT IN connector. The sig-
nal is routed to the filter out detector in the input sec-
tion.
604
RF Input/Output Test
Explanation and Suggestions
Measurement Point
HP-IB "MEAS:RFR:POWER?"
statement
DVM:FILTER_OUT_DET
N:\mkt\MANUALS\HP8924C\ALR\diagnos.fb

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8924e

Table of Contents