HP 8924C Assembly page 303

Mobile station test set
Hide thumbs Also See for 8924C:
Table of Contents

Advertisement

NOTE:
Procedure
N:\mkt\MANUALS\HP8924C\ALR\24e_pt.fb
CDMA Analyzer Relative Tuned Channel Power Level Accuracy Performance Test 32
Signal Generator
Set the frequency to 840 MHz.
Set the output level to 1 dBm.
Measuring Receiver
Set the display mode to LOG.
Set the measurement to Tuned RF Level.
Calibrate the power sensor.
The power sensor's calibration data must be entered into the measuring receiver and the sensor
module must be zeroed. Refer to the measuring receiver's and the sensor module's operating
manuals.
1. Press the PRESET key on the Test Set.
2. Access the CONFIGURE screen.
3. Set the RF Display field to Freq.
4. Set the (GEN)-(ANL) field to 0 MHz.
5. Change the Input Atten field to Hold and 0 dB.
6. Access the CDMA CALL CTRL screen.
7. Set the Avg Power field to Chan Power.
8. Turn Sector A Power to Off.
9. Set the RF Gen Freq to the values listed in PTR.
10. Measure each of the RF levels in the PTR with the HP 8924E and the Measuring
Receiver
(see "CDMA Analyzer Relative Tuned Channel Power Level Accuracy
Performance Test 32" on page
11. Use the following measurement equation and enter the result in the PTR.
A
= R
- R
meas
8924E
A
= Measured value
meas
R
= Power reading on HP 8924E
8924E
R
= Power Reading on Measuring Receiver
8902A
12. Change the Frequency on the Signal Generator, Measuring Receiver, and the
HP 8924E to each of the values in the PTR and repeat the above procedure.
404).
8902A
Chapter 6, HP 8924E Performance Verification
303