Chapter 6, HP 8924E Performance Verification
RF Gen FM Accuracy Performance Test 2
RF Gen FM Accuracy
Performance Test 2
Description
Setup
Measuring Receiver
236
The FM distortion of the RF generator is measured directly by the measuring
receiver. The Test Set's internal audio generator provides the modulation source.
INPUT 50
Make the following measuring receiver settings:
Reset the measuring receiver
300 Hz High-Pass Filter
3 kHz Low-Pass Filter
FM mode
FM de-emphasis off