HP 8924C Assembly page 269

Mobile station test set
Hide thumbs Also See for 8924C:
Table of Contents

Advertisement

Procedure
N:\mkt\MANUALS\HP8924C\ALR\24e_pt.fb
1. Press the PRESET key on the Test Set.
2. Access the ANALOG MES screen, and set the AF Anl In field to Audio In.
3. Set the Filter 2 field to >99 kHz LP, and the lower-right measurement field to
AF Freq.
4. Read the AF Freq on the Test Set and compare the reading to the limits shown in the
PTR
(see "AF Analyzer Frequency Accuracy at 400 kHz Performance Test 19" on
page
390).
Chapter 6, HP 8924E Performance Verification
AF Analyzer Frequency Accuracy at 400 kHz Performance Test 19
269