Procedure
N:\mkt\MANUALS\HP8924C\ALR\24e_pt.fb
1. Press the PRESET key on the Test Set.
2. Access the CONFIGURE screen.
3. Change the RF Display field to Freq.
4. Set the Output Port field to Dupl.
5. Select the ANALOG MEAS screen.
6. Set the AFGen1 level to Off.
7. Set the Test Set to the RF Generator frequencies and level shown in the PTR
Gen Harmonics Spectral Purity Performance Test 8" on page
the level of the second and third harmonics and the half-harmonics also shown in the
PTR.
8. Convert the harmonic levels to decibels below the fundamental (dBc) and compare the
computed levels to the limits shown in the PTR.
Chapter 6, HP 8924E Performance Verification
RF Gen Harmonics Spectral Purity Performance Test 8
(see "RF