HP 8924C Assembly page 217

Mobile station test set
Hide thumbs Also See for 8924C:
Table of Contents

Advertisement

CDMA GENERATOR CONTROL
Sector B
Sector A
Power
Power
dBm/BW
Pilot
Pilot
Walsh
0
dB
Sync
32
dB
Paging
1
dB
Traffic
Traffic
dB
OCNS
OCNS
-1.6928
-1.1435
dB
PN Offset
PN Offset
N:\mkt\MANUALS\HP8924C\ALR\24c_pt.fb
CDMA Generator Duplex Out Amplitude Level Accuracy Performance Test 28
9. Go to the CDMA CELLULAR MOBILE TRANSMITTER TEST screen.
10. Read the Avg Power (Average Power) and record it in the PTR.
11. Access the CDMA GENERATOR CONTROL screen.
12. Set the Sector B Power field to Off.
13. Access the CDMA CELLULAR MOBILE TRANSMITTER TEST screen.
14. Change the Avg Power measurement field to Chan Power (Channel Power).
15. Calibrate the Channel Power by selecting Calibrate in the Power Meas field.
16. Access the CDMA GENERATOR CONTROL screen.
17. Set the AWGN Power field to the values listed in the PTR for AWGN Power.
18. Read the Channel Power and record it in the PTR.
Set to Off
Set to value in PTR
AWGN
Walsh
RF Power
Eb/Nt
0
-75.0000
dB
dBm/BW
CDMA
Analog
dB
Config
dB
Chapter 5, HP 8924C Performance Verification
Select Chan Power
CDMA CCELLULAR MOBILE TRANSMITTER TEST
_ _ _ _
Meas Cntl
Read Channel Power
MS FER
Traffic
Sector A
To Screen
Data Mode
Power
CDMA
Data Type
Echo Delay
Analog
RF Power
-75.0000
dBm/BW
Config
Power Meas
Select Calibrate
217

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8924e