Appendix; Echo Curve - FlowLine EchoWave LG11 Series Manual

Guided wave liquid level transmitter
Hide thumbs Also See for EchoWave LG11 Series:
Table of Contents

Advertisement

Appendix

ECHO CURVE

This function displays the primary echo return(s) that the sensor is seeing graphically as well as the location
and amplitude of the return(s). It can be used to confirm the correct level reading by the sensor or to
troubleshoot any false signals. There is a two step process involving the creating and viewing of an echo
curve. Note: Always consult a Flowline
#1 - CREATE AN ECHO CURVE
To create an echo curve, use the following instructions:
1. Make sure the LG10/LG11 series is installed within the actual
application.
a. Echo curve will not provide any useful information when
placed outside of the application.
2. Press and hold the SCAN button for 1 second (LED will turn off),
then release.
3. Echo curve is completed.
a. To view, you must connect the sensor to WebCal
Note: LG10/LG11 series can store up to a total of three empty signal scans
and/or echo curve. Example: If an empty signal scan is active, EchoWave
will only store two echo curves. If an empty signal scan is not active, three
echo curves can be stores. Echo curves follow a first in first out logic.
#2 - VIEWING THE ECHO CURVE
The echo curve(s) scan can be viewed with the WebCal
the directions below:
1. Create an echo curve (see
instructions above).
2. Connect EchoWave
®
WebCal
software via Fob
3. Click on Diagnostics Tab on
the main WebCal
4. In the Select Signal Data pull
down, select Echo Curve.
5. The echo curve will be
displayed in the window.
|
32
®
representative for reviewing of the signal data.
®
to
®
screen
®
.
®
®
software (version 6.5 and greater). To view, follow
Section Seven
MN300840 REV A5