Specification Information

Transportation and
storage conditions
Operating conditions
Standards for
electrical equipment
Co m bi n a ti o n B l an k et / Fl u i d W ar m er
Model
Intended use
TriMode cavity—warming blankets, irrigation or injection fluids
Weight
TriMode cavity: 30 1-liter
bottles or 24 1-liter bags, or
Storage cavity
10–14 blankets (5.4 ft
capacity
Blanket cavity: 30–40
blankets (13 ft
Temperature
range
Clearance
requirements
Ambient temperature range of -40°F to +159°F (-40°C to +70°C).
Relative humidity range of 10% to 95%, non-condensing.
Atmospheric pressure range of 7.25 psi to 15.37 psi (50kPa to 106kPa).
The warmer must acclimate to the room temperature it will be placed in—24
hours is recommended.
The recommended environmental temperature range is 60°F to 90°F
(15°C to 32°C).
The recommended relative humidity is above 20%, non-condensing.
Medical equipment listed by Underwriters Laboratories with respect to
electrical shock, fire, and mechanical hazards only, in accordance with UL
61010-1 and CAN/CSA C22.2 No. 61010-1.
Grounding reliability can only be achieved when the appliance is
connected to an equivalent receptacle marked "Hospital Grade."
EN61326-1; EN61000-3-2; EN61000-3-3
IP-XO ordinary
Ope r at or ' s Man u a l
MN -3 9 9 30
P-2147
Bottom cavity—warming blankets
Net: 457 lb (207 kg)
Ship: 528 lb (239 kg)
3
)
3
)
Irrigation fluids (IRR):
90°F to 150°F (32°C to 66°C)
Injection fluids (INJ):
90°F to 104°F (32°C to 40°C)
Blanket: 90°F to 160°F (32°C to 71°C)
4" (102mm) from rear
1" (25mm) from top
2" (51mm) from sides
1/2" (13mm) from bottom
Re v 2
4 / 2 0
S
PECIFICATIONS
P-2150
Net: 336 lb (152 kg)
Ship: 407 lb (185 kg)
TriMode cavity: 24 1-liter
bottles or 20 1-liter bags, or
3
9–12 blankets (3.9 ft
)
Blanket cavity: 24–30
3
blankets (9.4 ft
)
2" (102mm) from rear
1" (25mm) from top
1" (51mm) from sides
NOTE:
See topic Dimension
Drawings for P-2150 Built-In
requirements.
E471516
Protective Earth
Ground Symbol
13