Guidance and Manufacturer's Declaration
Electromagnetic
emissions
Co m bi n a ti o n B l an k et / Fl u i d W ar m er
The warmer requires special precautions regarding EMC (Electromagnetic
Compatibility) and needs to be installed and put into service according to the
EMC information provided in the accompanying documents.
Portable and mobile RF communications equipment can affect medical electrical
equipment.
A risk of increased emissions or decreased immunity may result if the power
cord is altered or a manufacturer supplied power cord is not used.
The warmer should not be used adjacent to or stacked with other equipment.
The essential performance of the warmer is to not exceed an internal
temperature of 160ºF (71ºC) for blanket warmers or 150ºF (66ºC) for fluid
warmers.
The warmer is intended for use in the electromagnetic environment specified
below.
Emission test
Compliance
RF emissions;
Group 1
CISPR 11
RF emissions;
Class B
CISPR 11
Harmonic
N/A
emissions; IEC
61000-3-2
Voltage
N/A
fluctuations/Flicker
emissions; IEC
61000-3-3
Ope r at or ' s Man u a l
MN -3 9 9 30
E
LECTRICAL
Electromagnetic environment - guidance
The warmer uses RF energy only for internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
The warmer is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Re v 2
4 / 2 0
I
NFORMATION
Continued on next page
53