How To Operate The Blanket Warmer (Bottom Cavity) - Pedigo P-2150 Operator's Manual

Combination blanket/fluid warmer
Table of Contents

Advertisement

O
PERATION
How to Operate the Blanket Warmer
(Bottom Cavity)
Before you begin
Procedure
During the warming
process
Result
30
The warmer must be connected to electric power and turned on.
WARNING:
Do not operate the warmer in the presence of a flammable
anesthetic mixture with air, oxygen, or nitrous oxide; in
oxygen-enriched environments; or in any other potentially
explosive environment.
WARNING:
Make sure the blanket support assembly and shelves are
installed to prevent the blankets from being scorched or
discolored.
Do not overload the cavity.
NOTICE
Allow 1" (25mm) between the blankets and the interior walls.
To operate the warmer, do the following.
Step
Action
Set
1.
the temperature using the arrow buttons
NOTE:
(32°C – 71°C).
2.
Open
the door and load the dry, cotton blankets into the warmer.
Close
the door.
NOTICE
3.
Press
the temperature recall button
temperature.
The measured cavity temperature will display for five seconds. Then, the
set-point temperature will display.
Rotate
4.
the blankets daily from the bottom to the top of the load to ensure
equal heating and to prevent discoloration.
The blankets are now warming.
Co m b i n at i on B l an k e t/ F l u id W a rm e r
Personal injury hazard.
Personal injury hazard.
The temperature set-point range is 90°F – 160°F
Do not warm items containing plastic, rubber, metal
snaps, studs, hooks, etc.
to view the measured cavity
Op er a to r ' s Ma n u al
.
M N- 39 9 3 0
Rev 2
4/ 2 0